BME Bergmann

Bergmann Messgeräte Entwicklung KG

Leysop EM510, 10mm Aperture KD*P Pockels Cell

This oscilloscope trace shows the measurement of the optical transmission in
a single-pass setup measured with a photodiode of risetime 150psec.

Parameters: 1064nm, 10.3kV, time difference of On/Off triggers 4.2nsec.

Risetime is 2.0ns, falltime is 1.7nsec

For pulse-picking applications, the driver can be operated in “overdrive” mode,
meaning that the voltage is increased while the time difference between ON/OFF triggers is decreased.

The oscilloscope trace below shows the measurement of the optical transmission in
a single-pass setup measured with a photodiode of risetime 150psec.

Parameters: 1064nm, 12kV, time difference of On/Off triggers 2.3nsec.

Risetime is 1.7ns, falltime is 1.9nsec